Fiber optic, photonic & electronic test & measurement
FiberLabs

C-band amplifier (EDFA, Bench Top)

Bench-top C-band erbium-doped fiber amplifiers in low-noise pre-amplifier, high-output booster and gain-flattened WDM variants. FiberLabs notes the design is not application specific and is also used for fiber sensing and fiber lasers.

C-band amplifier (EDFA, Bench Top)

Technical overview

The AMP-FL8013-CB is the low-noise-figure pre-amplifier line (NF <= 5 dB, output >= +13 to +19 dBm) and the AMP-FL8011-CB is the booster line (output up to >= +22 dBm, gain up to >= 40 dB, NF <= 6 dB). Both are specified 1530 to 1560 nm at a 1550 nm nominal wavelength, with the product header quoting a 1525 to 1570 nm range. The -W gain-flattened models hold <= 3 dB flatness with 1530 to 1560 nm signals at 0 dBm total input. All ship in an 88 x 260 x 350 mm chassis on G.652 SMF with FC/PC connectors.

Features

  • Proven in Various Deployment Conditions (from R&D to Actual Network)
  • Turn-Key Operation
  • RS-232C and GP-IB

Specifications

Specifications
ParameterValue
Product numbersAMP-FL8011-CB-19 / -22 / -22-W; AMP-FL8013-CB-13 / -16 / -19 / -19-W
Operating wavelength (spec table)1530 to 1560 nm
Operating wavelength (product header)1525 to 1570 nm
Nominal wavelength1550 nm
Output power (input 0 dBm at nominal wavelength)>= +13 / >= +16 / >= +19 / >= +22 by grade dBm
Gain (input -30 dBm at nominal wavelength)>= 35 (8011-CB-19) / >= 40 (8011-CB-22) / Typ. 35 (8011-CB-22-W) / >= 25 (8013-CB-13) / >= 30 (8013-CB-16, -19) / Typ. 30 (8013-CB-19-W) dB
Noise figure (input -30 dBm at nominal wavelength)<= 6 (8011-CB series) / <= 5 (8013-CB series) dB
Gain flatness (-W models)<= 3 (input signals 1530 to 1560 nm, total power 0 dBm) dB
Input/output fiberSMF, G.652 compliant (e.g. SMF28)
Standard connectorFC/PC
Input/output isolatorBuilt In
Standard dimensions (H x W x D)88 x 260 x 350 mm
Safety functionRemote Interlock
Control modeACC (Auto Current Control) and ALC (Auto Light Control)
Remote interfaceRS-232C / IEEE488.1 (GP-IB)
Operating temperature0 to 40 degC
Storage temperature-10 to 60 degC
Power supplyAC 100 to 240 V (50/60 Hz)
Specifications are reproduced from the manufacturer's own published material on the date shown and are subject to change. Confirm at order.

Applications

  • Laboratory Use
  • Test and Measurement
  • Integration in System
Manufacturer datasheet (PDF)

You may also like

Need help specifying this?

Our application engineers will confirm specifications with the manufacturer and quote for your configuration.

Request Info